Aure Chemical

MES vs MOPS Running Buffer for Bis-Tris Protein Gels

Persistent supply for specialized needs

MES vs MOPS Running Buffer for Bis-Tris Protein Gels


For compatible Bis-Tris protein gels, MES SDS running buffer is generally the starting choice for smaller proteins, while MOPS SDS running buffer is commonly used for medium-sized and larger proteins. The useful separation range also depends on gel percentage, gradient and format. Choose the gel and running buffer together, using the manufacturer's migration chart for the region that matters to your sample.[3]

The comparison below uses NuPAGE Bis-Tris denaturing SDS-PAGE as a defined example. It covers the choice between complete MES SDS and MOPS SDS formulations, followed by the checks needed when sourcing MES, MOPS and Tris base as raw materials.

Check the exact gel system first

Read the gel label and its current instructions before selecting a running buffer. NuPAGE Bis-Tris and Bolt Bis-Tris Plus systems both have recommended MES SDS and MOPS SDS buffers, but their sample buffers, run settings and other instructions must be matched to the relevant product.[3] A shared chemical name on two bottles does not establish compatibility with every gel.

NativePAGE Bis-Tris uses a different reagent system; Tris-glycine and Tris-acetate methods also have their own running-buffer formulations.[1] Keep the complete method together when making a change. Also distinguish Bis-Tris, the chemical used in the gel chemistry, from the Tris base listed in the running-buffer recipe. They are different compounds.

For the wider choice of buffer chemistry and working conditions, see the Aure Chemical biological buffer selection guide.

Choose for the bands you need to separate

A broad molecular-weight range on a product page is useful for screening options. For a close separation, look at the spacing around your target on the chart for the actual gel percentage. A target that sits within the nominal range may still be poorly separated from a neighbouring band.

MES SDS and MOPS SDS: selection checks for compatible Bis-Tris protein gels
Decision pointMES SDSMOPS SDSWhat to check
Smaller target proteinsUsually the first option to evaluate in this system.[3]Compare the chart before choosing it for a low-molecular-weight target.Spacing around the target, gel percentage and ladder coverage.
Medium-sized or larger target proteinsCheck whether the target region has enough separation on the chosen gel.Commonly selected for this region in compatible Bis-Tris systems.[3]The specific fixed-percentage or gradient gel and its migration chart.
Two closely spaced bandsEvaluate the local band spacing.Evaluate the same region under matched conditions.Whether the change improves separation of those bands, rather than simply moving them.
Smiling, tailing or weak bandsRecord the problem before attributing it to the choice of MES or MOPS.Review sample preparation, loading, buffer preparation, gel condition and the equipment instructions.

There is no single kDa cutoff that separates all MES applications from all MOPS applications. If comparing the two, keep the gel type, sample preparation, amount loaded and ladder consistent. Assess the separation you need; a change in migration does not mean that the protein's true molecular mass has changed.

Compare the complete 1X running-buffer formulations

MES SDS and MOPS SDS running buffers contain several components. Preparing MES or MOPS alone in water will not reproduce either formulation. The following working concentrations are listed for the NuPAGE Bis-Tris method.[1]

Published 1X compositions for NuPAGE MES SDS and MOPS SDS running buffers
Component or conditionMES SDS, 1XMOPS SDS, 1XPreparation check
MES or MOPS50 mM MES50 mM MOPSMatch the specified chemical form to the preparation route.
Tris base50 mM50 mMTris base and Tris-HCl are different raw materials.
SDS0.1% (w/v)0.1% (w/v)Use the concentration in the final working volume.
EDTA1 mM1 mMConfirm the chemical form and hydration state before calculating a mass.
Listed working pH7.37.7These values refer to the 1X running solutions.

The NuPAGE quick reference specifies that the pH values apply to the 1X solutions and instructs users not to adjust these recipes with acid or base.[2] That instruction belongs to these formulations. It should not be carried over to unrelated MES or MOPS buffers. An unexpected reading calls for a check of the recipe, dilution, reagent identities and measurement conditions before changing the formulation.

A 20X label describes concentration, not pH. For 100 mL of a 1X working buffer, dilute 5 mL of the appropriate 20X concentrate with water to a final volume of 100 mL. Do not assume that the concentrate has the same measured pH as the working solution.

Match the sample buffer and reducing conditions

Running buffer, sample buffer and transfer buffer have separate jobs. The cited NuPAGE method uses an LDS sample buffer.[1] A Laemmli-style loading buffer should therefore be assessed against the chosen method before substitution. SDS and LDS are detergents; neither replaces the reducing agent used to break disulfide bonds.

For reduced samples, the NuPAGE instructions specify a separate sample reducing agent and an antioxidant in the running buffer. Follow the directions for the tank in use, including chamber placement and quantities.[2] Treat antioxidant use and sample reduction as separate parts of the procedure.

These are denaturing electrophoresis methods. A near-neutral gel chemistry does not establish that native folding or enzyme activity will be preserved. If the next step is a Western blot, check the transfer-buffer instructions separately rather than reusing the running-buffer composition.

Distinguish protein and RNA MOPS buffers

The RNA agarose-gel protocol cited below uses a MOPS running buffer containing sodium acetate and EDTA. It is a different complete formulation from the MOPS, Tris base, SDS and EDTA protein running buffer described here.[4] When requesting or preparing a MOPS buffer, identify the application and full recipe.

Specify MES, MOPS and Tris base as raw materials

For preparation from raw materials, distinguish MES free acid (CAS 4432-31-9), MOPS free acid (CAS 1132-61-2) and Tris base (CAS 77-86-1) on the purchasing specification. These products are ingredients for formulation; their names alone do not describe a finished SDS running buffer.

Free acids and sodium salts require different formulation calculations. A sodium salt changes both formula weight and counterion contribution, so it should not be substituted gram-for-gram into a free-acid recipe. Check the assay basis and water or hydrate content as well, avoiding a second moisture correction if it is already included in the calculation.

Review the available specification and lot documentation against the receiving laboratory's requirements. Chemical identity and assay alone do not establish the band resolution of a finished running buffer; any electrophoresis-performance requirement should be defined and evaluated for the intended method.

For sourcing through Aure Chemical, send the required chemical form, quantity, packaging, destination and documentation requirements to sales@aurechem.com. Include the intended formulation when requesting a raw-material specification.

Technical references

  1. Thermo Fisher Scientific: Protein Electrophoresis Buffers and Reagents.

  2. Thermo Fisher Scientific: NuPAGE Bis-Tris Mini Gels Quick Reference, MAN0007891, Rev. D.

  3. Thermo Fisher Scientific: NuPAGE Bis-Tris and Bolt Bis-Tris Plus Gels.

  4. Thermo Fisher Scientific: Agarose Gel Electrophoresis of RNA.

>